Crowcon Xgard ASU Air Sampling Unit
The Air Sampling Unit (ASU) is designed to be used with Crowcon’s Xgard
detectors to monitor potentially hazardous gases within ventilation ducts.
The ASU uses the Venturi principle, which states that the airflow rate within
the duct must be between 5 and 20 metres per second.
Installation
The Xgard gas detector must be fitted to the ASU before being assembled onto
the duct. The ASU is not compatible with the Xgard Type 4 high-temperature
gas detector.
- To avoid the adverse effects of turbulence, the ASU should be installed
in the centre of a straight section of the duct, which has a length of
at least six times its width.
- Check the airflow direction in the air duct system.
- Apply the drilling template onto the centre line of the air duct
parallel to the airflow. Check that the directional arrow on the
drilling template is the same as the direction of the airflow in the
duct.
- Punch the two centres indicated for the drilling template using a centre
punch.
- Drill out to 25mm (1”) diameter. Remove burrs and sharp edges around the
two holes.
- Remove the ASU cover.
- Fit the sample tube elbows with rubber washers inside and outside the
sampling chamber. Fit the securing nuts and tighten. Ensure that the
black sample tube holes face 180º from the sample chamber and the white
sample tube chamfer faces away from the sample chamber.
- Check the thickness of the air duct metalwork and fit the black rubber
grommets into the two duct holes drilled for point 5 of these
instructions.
- Fit the ASU probes through the two grommets, check that the inlet and
outlet probes are correctly installed, and push until the unit has been
sealed.
- Mark the six ASU plate fixing holes and drill 5mm (1/4”) holes for
securing the plate. Secure the plate to the duct using suitable bolts.
- Connect the detector cable as instructed in the Xgard manual.
- Check that there is no air leakage into the sampling chamber from
outside the duct through the probe joints, grommets or sampling unit
cover gasket.
Testing and Commissioning
Calibrate the detector according to its instructions. Test the operation of
the ASU by introducing gas into the duct system.
Warning: it is essential to ensure that procedures are in place to prevent
people and equipment from being affected by any test gas introduced into the
ventilation system.
Operation and maintenance
The ASU needs no specific maintenance; however, the Xgard gas detector must
be maintained according to its instructions.